Development of a Flat-Panel and Resistor-Type Photomultiplier Tube System for High Position-Resolution Two-Dimensional Neutron Detector
Setsuo SATOH
High Energy Accelerator Research Organization (KEK), 1-1 Oho, Tsukuba, Ibaraki 305-0801, Japan
(Received 28 May 2017 / Accepted 21 December 2017 / Published 22 May 2018)

Abstract

A flat-panel and resistor-type photomultiplier tube (FRP) system, which is a two-dimensional position-sensitive neutron detector, has been developed. The FRP detector consists of a scintillator, a multi-anode photomultiplier tube (MA-PMT), a resistor network board, and an FRP amplifier and a readout circuit which is similar to that for the old resistor type photomultiplier tube (RPMT) system. Therefore, the FRP system uses the same readout circuit of the RPMT system. Two types of FRP detector are available. One is the FRP4 detector, which consists of four MA-PMTs with a detection area of 10-cm square. The position resolution due to irradiation of neutron beam with approximately 0.5-mm diameter is 1.26 mm in the X axis and 1.61 mm in the Y axis. The other one is the FRP1, which consists of one MA-PMT with a detection area of 5-cm square. The position resolution due to irradiation of a neutron beam with approximately 0.5-mm diameter is 0.83 mm in the X axis and 0.89 mm in the Y axis. The FRP detector can be used for high position-resolution and transmission experiments of a compact neutron source.


Keywords

FRP system, RPMT system, two-dimensional neutron detector, high position-resolution neutron detector, compact neutron source

DOI: 10.1585/pfr.13.2405056


References

  • [1] S. Satoh et al., IPS08 Proceedings of the International Symposium on Pulsed Neutron and Muon Sciences at J-PARC 2008, DOI:10.1016/j.nima.2008.11.054.
  • [2] S. Satoh, “Development of readout systems for neutron detectors,” Hamon 15(1), 78-81 (2005), in Japanese.
  • [3] Hamamatsu Photoniccs, R3292 Datasheet, TPMH1238E01 Sept. 1998.